Abstract
This paper presents a portable and contactless microfluidic flowmeter based on the Nuclear Magnetic Resonance (NMR) principle, which can measure flow rates up to 2.7 ml/min with \pm 16\ lmin (±0.6% full-scale) error. It consists of a millimeter-size RF coil wound around a microfluidic channel, custom readout electronics and a 0.5 T permanent magnet with 50 ppm inhomogeneity. To our knowledge, this is the first reported low-field NMR microfluidic flowmeter.
Original language | English |
---|---|
Title of host publication | 21st International Conference on Solid-State Sensors, Actuators and Microsystems, TRANSDUCERS 2021 |
Publisher | IEEE |
Pages | 1020-1023 |
Number of pages | 4 |
ISBN (Electronic) | 978-1-6654-1267-4 |
ISBN (Print) | 978-1-6654-4845-1 |
DOIs | |
Publication status | Published - 2021 |
Event | 2021 21st International Conference on Solid-State Sensors, Actuators and Microsystems (Transducers) - Virtual at Orlando, United States Duration: 20 Jun 2021 → 24 Jun 2021 Conference number: 21st |
Publication series
Name | 21st International Conference on Solid-State Sensors, Actuators and Microsystems, TRANSDUCERS 2021 |
---|
Conference
Conference | 2021 21st International Conference on Solid-State Sensors, Actuators and Microsystems (Transducers) |
---|---|
Country/Territory | United States |
City | Virtual at Orlando |
Period | 20/06/21 → 24/06/21 |
Bibliographical note
Green Open Access added to TU Delft Institutional Repository 'You share, we take care!' - Taverne project https://www.openaccess.nl/en/you-share-we-take-careOtherwise as indicated in the copyright section: the publisher is the copyright holder of this work and the author uses the Dutch legislation to make this work public.
Keywords
- NMR
- analog front-end
- flowmeter
- low noise transceiver
- microfluidics