Application of the ESHO-QA guidelines for determining the performance of the LCA superficial hyperthermia heating system

Carolina Carrapiço-Seabra*, Mattia De Lazzari, Abdelali Ameziane, Gerard C. van Rhoon, Hana Dobšícek Trefná, Sergio Curto

*Corresponding author for this work

Research output: Contribution to journalArticleScientificpeer-review

29 Downloads (Pure)

Abstract

Purpose: This study aimed to assess the quality of the lucite cone applicator (LCA), the standard applicator for superficial hyperthermia at the Erasmus MC Cancer Institute, using the most recent quality assurance guidelines, thus verifying their feasibility. Materials and methods: The assessment was conducted on each of the six LCAs available for clinical treatments. The temperature distribution was evaluated using an infrared camera across different layers of a fat-muscle mimicking phantom. The maximum temperature increase, thermal effective penetration depth (TEPD), and thermal effective field size (TEFS) were used as quality metrics. The experimental results were validated through comparison with simulated results, using a canonical phantom model and a realistic phantom model segmented from CT imaging. Results: A maximum temperature increase above 6 °C at 2 cm depth in the fat-muscle phantom for all the experiments was found. A mean negative difference between simulated and experimental data was of 1.3 °C when using the canonical phantom model. This value decreased to a mean negative difference of 0.4 °C when using the realistic model. Simulated and measured TEPD showed good agreement for both in silico scenarios, while discrepancies were present for TEFS. Conclusions: The LCAs passed all QA guidelines requirements for superficial hyperthermia delivery when used singularly or in an array configuration. A further characterization of parameters such as antenna efficiency and heat transfer coefficients would be beneficial for translating experimental results to simulated values. Implementing the QA guidelines was time-consuming and demanding, requiring careful preparation and correct setup of antenna elements.

Original languageEnglish
Article number2272578
Number of pages13
JournalInternational Journal of Hyperthermia
Volume40
Issue number1
DOIs
Publication statusPublished - 2023

Keywords

  • Hyperthermia
  • LCA
  • quality assurance (QA)
  • superficial applicators

Fingerprint

Dive into the research topics of 'Application of the ESHO-QA guidelines for determining the performance of the LCA superficial hyperthermia heating system'. Together they form a unique fingerprint.