T1 - Design and Proof-of-Concept of a Matrix Transducer Array for Clamp-on Ultrasonic Flow Measurements
AU - Massaad, Jack
AU - Van Neer, Paul L.M.J.
AU - Van Willigen, Douwe M.
AU - Noothout, Emile C.
AU - De Jong, Nicolaas
AU - Pertijs, Michiel A.P.
AU - Verweij, Martin D.
PY - 2022
Y1 - 2022
N2 - Common clamp-on ultrasonic flow meters consist of two single-element transducers placed on the pipe wall. Flow speed is measured noninvasively, i.e., without interrupting the flow and without perforating the pipe wall, which also minimizes safety risks and avoids pressure drops inside the pipe. However, before metering, the transducers have to be carefully positioned along the pipe axis to correctly align the acoustic beams and obtain a well-calibrated flowmeter. This process is done manually, is dependent on the properties of the pipe and the liquid, does not account for pipe imperfections, and becomes troublesome on pipelines with an intricate shape. Matrix transducer arrays are suitable to dynamically steer acoustic beams and realize self-alignment upon reception, without user input. In this work, the design of a broadband 37×17 matrix array (center frequency of 1 MHz) to perform clamp-on ultrasonic flow measurements over a wide range of liquids (c=1000-2000m/s, α≤1 dB/MHz · cm) and pipe sizes is presented. Three critical aspects were assessed: efficiency, electronic beam steering, and wave mode conversion in the pipe wall. A prototype of a proof-of-concept flowmeter consisting of two 36-element linear arrays (center frequency of 1.1 MHz) was fabricated and placed on a 1-mm-thick, 40-mm inner diameter stainless steel pipe in a custom-made flow loop filled with water. At resonance, simulated and measured efficiencies in water of the linear arrays compared well: 0.88 and 0.81 kPa/V, respectively. Mean flow measurements were achieved by electronic beam steering of the acoustic beams and using both compressional and shear waves generated in the pipe wall. Correlation coefficients of R2>0.99 between measured and reference flow speeds were obtained, thus showing the operational concept of an array-based clamp-on ultrasonic flowmeter.
AB - Common clamp-on ultrasonic flow meters consist of two single-element transducers placed on the pipe wall. Flow speed is measured noninvasively, i.e., without interrupting the flow and without perforating the pipe wall, which also minimizes safety risks and avoids pressure drops inside the pipe. However, before metering, the transducers have to be carefully positioned along the pipe axis to correctly align the acoustic beams and obtain a well-calibrated flowmeter. This process is done manually, is dependent on the properties of the pipe and the liquid, does not account for pipe imperfections, and becomes troublesome on pipelines with an intricate shape. Matrix transducer arrays are suitable to dynamically steer acoustic beams and realize self-alignment upon reception, without user input. In this work, the design of a broadband 37×17 matrix array (center frequency of 1 MHz) to perform clamp-on ultrasonic flow measurements over a wide range of liquids (c=1000-2000m/s, α≤1 dB/MHz · cm) and pipe sizes is presented. Three critical aspects were assessed: efficiency, electronic beam steering, and wave mode conversion in the pipe wall. A prototype of a proof-of-concept flowmeter consisting of two 36-element linear arrays (center frequency of 1.1 MHz) was fabricated and placed on a 1-mm-thick, 40-mm inner diameter stainless steel pipe in a custom-made flow loop filled with water. At resonance, simulated and measured efficiencies in water of the linear arrays compared well: 0.88 and 0.81 kPa/V, respectively. Mean flow measurements were achieved by electronic beam steering of the acoustic beams and using both compressional and shear waves generated in the pipe wall. Correlation coefficients of R2>0.99 between measured and reference flow speeds were obtained, thus showing the operational concept of an array-based clamp-on ultrasonic flowmeter.
KW - Acoustics
KW - Beam steering
KW - beam steering
KW - Biomedical measurement
KW - clamp-on flow meter
KW - Guided waves
KW - Liquids
KW - Meters
KW - transducer design
KW - Ultrasonic transducer arrays
KW - Ultrasonic variables measurement
KW - ultrasound flow meter
UR - http://www.scopus.com/inward/record.url?scp=85133786752&partnerID=8YFLogxK
U2 - 10.1109/TUFFC.2022.3186170
DO - 10.1109/TUFFC.2022.3186170
M3 - Article
AN - SCOPUS:85133786752
SN - 0885-3010
VL - 69
SP - 2555
EP - 2568
JO - IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control
JF - IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control
IS - 8