TY - JOUR
T1 - Directional band gap phononic structures for attenuating crosstalk in clamp-on ultrasonic flowmeters
AU - Valappil, Sabiju Valiya
AU - Aragón, Alejandro M.
AU - Goosen, Johannes F.L.
PY - 2025
Y1 - 2025
N2 - Clamp-on ultrasonic flowmeters suffer from crosstalk—i.e., measurement errors due to the interference of signals generated in solid regions and solid–fluid interfaces with the required signal from the fluid. Although several approaches have been proposed to alleviate crosstalk, they only work in specific ranges of flow rates and pipe diameters, and some also introduce additional issues. We propose a novel clamp-on system design where the transmitting and receiving wedges are embedded with directional noise filtering mechanisms based on phononic crystals (PnCs) possessing directional band gaps (DBGs). PnCs are artificial materials consisting of periodic structures arrayed in a matrix medium exhibiting band gaps – i.e., frequency ranges where waves are attenuated – due to Bragg scattering. DBGs enable PnCs to propagate waves in specific directions while suppressing them in other directions. By guiding the input signal through the transmitting wedge to the wall, we minimize the generation of noise signals due to secondary reflections within the wedge. Similarly, by using the directionality of the DBG PnC in the receiver, we limit the effects of noise signals (that arrive in different directions) in the receiver. We numerically verify the DBG PnC embedded wedges’ performance by comparing wave propagation aspects of the PnC embedded clamp-on system with a standard clamp-on device. To that end, we develop accurate wave propagation models based on the Discontinuous Galerkin finite element method. By incorporating DBG PnCs into the wedges, we obtain about 20 dB increase in the signal-to-noise ratio compared to the clamp-on system with standard wedges.
AB - Clamp-on ultrasonic flowmeters suffer from crosstalk—i.e., measurement errors due to the interference of signals generated in solid regions and solid–fluid interfaces with the required signal from the fluid. Although several approaches have been proposed to alleviate crosstalk, they only work in specific ranges of flow rates and pipe diameters, and some also introduce additional issues. We propose a novel clamp-on system design where the transmitting and receiving wedges are embedded with directional noise filtering mechanisms based on phononic crystals (PnCs) possessing directional band gaps (DBGs). PnCs are artificial materials consisting of periodic structures arrayed in a matrix medium exhibiting band gaps – i.e., frequency ranges where waves are attenuated – due to Bragg scattering. DBGs enable PnCs to propagate waves in specific directions while suppressing them in other directions. By guiding the input signal through the transmitting wedge to the wall, we minimize the generation of noise signals due to secondary reflections within the wedge. Similarly, by using the directionality of the DBG PnC in the receiver, we limit the effects of noise signals (that arrive in different directions) in the receiver. We numerically verify the DBG PnC embedded wedges’ performance by comparing wave propagation aspects of the PnC embedded clamp-on system with a standard clamp-on device. To that end, we develop accurate wave propagation models based on the Discontinuous Galerkin finite element method. By incorporating DBG PnCs into the wedges, we obtain about 20 dB increase in the signal-to-noise ratio compared to the clamp-on system with standard wedges.
KW - Directional band gap
KW - Flow measurement
KW - Noise reduction
KW - Phononic crystals
KW - Ultrasonic transducer
KW - Wave guiding
UR - http://www.scopus.com/inward/record.url?scp=85211367981&partnerID=8YFLogxK
U2 - 10.1016/j.ymssp.2024.112173
DO - 10.1016/j.ymssp.2024.112173
M3 - Article
AN - SCOPUS:85211367981
SN - 0888-3270
VL - 224
JO - Mechanical Systems and Signal Processing
JF - Mechanical Systems and Signal Processing
M1 - 112173
ER -