Imaging performance of a LaBr3:Ce scintillation detector for photon counting x-ray computed tomography: Simulation study

Katsuyuki Taguchi*, Dennis R. Schaart, Marlies C. Goorden, Scott S. Hsieh

*Corresponding author for this work

Research output: Contribution to journalArticleScientificpeer-review

Abstract

Background
Photon counting detectors (PCDs) for x-ray computed tomography (CT) are the future of CT imaging. At present, semiconductor-based PCDs such as cadmium telluride (CdTe), cadmium zinc telluride, and silicon have been either used or investigated for clinical PCD CT. Unfortunately, all of them have the same major challenges, namely high cost and limited spectral signal-to-noise ratio (SNR). Recent studies showed that some high-quality scintillators, such as lanthanum bromide doped with cerium (LaBr3:Ce), are less expensive and almost as fast as CdTe.

Purpose
The objective of this study is to assess the performance of a LaBr3:Ce PCD for clinical x-ray CT.

Methods
We performed Monte Carlo simulations and compared the performance of 3 mm thick LaBr3:Ce and 2 mm thick CdTe for PCD CT with x-rays at 120 kVp and 20–1000 mA. The two PCDs were operated with either a threshold–subtract (TS) counting scheme or a direct energy binning (DB) counting scheme. The performance was assessed in terms of the accuracy of registered spectra, counting capability, and count-rate-dependent spectral imaging-task performance, for conventional CT imaging, water–bone material decomposition, and K-edge imaging with tungsten as the K-edge material. The performance for these imaging-tasks was quantified by nCRLB, that is, the Cramér–Rao lower bound on the variance of basis line-integral estimation, normalized by the corresponding value of CdTe at 20 mA.

Results
The spectrum recorded by CdTe was distorted significantly due to charge sharing, whereas the spectra recorded by LaBr3:Ce better matched the incident spectrum. The dead time, estimated by fitting a paralyzable detector model to the count-rate curves, was 20.7, 15.0, 37.2, and 13.0 ns for CdTe with TS, CdTe with DB, LaBr3:Ce with TS, and LaBr3:Ce with DB, respectively. Conventional CT imaging showed an adverse effect of reduced geometrical efficiency due to optical reflectors in LaBr3:Ce PCD. The nCRLBs (a lower value indicates a better SNR) for CdTe with TS, CdTe with DB, LaBr3:Ce with TS, LaBr3:Ce with DB, and the ideal PCD, were 1.00 ± 0.01, 1.00 ± 0.01, 1.18 ± 0.02, 1.18 ± 0.02, and 0.79 ± 0.01, respectively, at 20 mA. The nCRLBs for water–bone material decomposition, in the same order, were 1.00 ± 0.02, 1.00 ± 0.02, 0.85 ± 0.02, 0.85 ± 0.02, and 0.24 ± 0.02, respectively, at 20 mA; and 0.98 ± 0.02, 0.98 ± 0.02, 1.09 ± 0.02, 0.83 ± 0.02, and 0.24 ± 0.02, respectively, at 1000 mA. Finally, the nCRLBs for K-edge imaging, the most demanding task among the five, were 1.00 ± 0.02, 1.00 ± 0.02, 0.55 ± 0.02, 0.55 ± 0.02, and 0.13 ± 0.02, respectively, at 20 mA; and 2.45 ± 0.02, 2.29 ± 0.02, 3.12 ± 0.02, 2.11 ± 0.02, and 0.13 ± 0.02, respectively, at 1,000 mA.

Conclusion
The Monte Carlo simulations showed that, compared to CdTe with either TS or DB, LaBr3:Ce with DB provided more accurate spectra, comparable or better counting capability, and superior spectral imaging-task performances, that is, water–bone material decomposition and K-edge imaging. CdTe had a better performance than LaBr3:Ce for the conventional CT imaging task due to its higher geometrical efficiency. LaBr3:Ce PCD with DB scheme may be an excellent alternative option for CdTe PCD.
Original languageEnglish
Pages (from-to)158-170
Number of pages13
JournalMedical Physics
Volume52
Issue number1
DOIs
Publication statusPublished - 2025

Bibliographical note

Green Open Access added to TU Delft Institutional Repository 'You share, we take care!' - Taverne project https://www.openaccess.nl/en/you-share-we-take-care
Otherwise as indicated in the copyright section: the publisher is the copyright holder of this work and the author uses the Dutch legislation to make this work public.

Keywords

  • computed tomography
  • photon counting detector
  • scintillator

Fingerprint

Dive into the research topics of 'Imaging performance of a LaBr3:Ce scintillation detector for photon counting x-ray computed tomography: Simulation study'. Together they form a unique fingerprint.

Cite this