TY - JOUR
T1 - On the durability of surgical masks after simulated handling and wear
AU - Varanges, Vincent
AU - Caglar, Baris
AU - Lebaupin, Yann
AU - Batt, Till
AU - He, Weidong
AU - Wang, Jing
AU - Rossi, René M.
AU - Richner, Gilles
AU - Delaloye, Jean Romain
AU - Michaud, Véronique
PY - 2022
Y1 - 2022
N2 - After the spread of COVID-19, surgical masks became highly recommended to the public. They tend to be handled and used multiple times, which may impact their performance. To evaluate this risk, surgical masks of Type IIR were submitted to four simulated treatments: folding, ageing with artificial saliva or sweat and washing cycles. The air permeability, mechanical integrity, electrostatic potential, and filtration efficiency (FE) of the masks were measured to quantify possible degradation. Overall, air permeability and mechanical integrity were not affected, except after washing, which slightly degraded the filtering layers. Electrostatic potential and FE showed a strong correlation, highlighting the role of electrostatic charges on small particle filtration. A slight decrease in FE for 100 nm particles was found, from 74.4% for the reference masks to 70.6% for the mask treated in saliva for 8 h. A strong effect was observed for washed masks, resulting in FE of 46.9% (± 9.5%), comparable to that of a control group with no electrostatic charges. A dry store and reuse strategy could thus be envisaged for the public if safety in terms of viral and bacterial charge is ensured, whereas washing strongly impacts FE and is not recommended.
AB - After the spread of COVID-19, surgical masks became highly recommended to the public. They tend to be handled and used multiple times, which may impact their performance. To evaluate this risk, surgical masks of Type IIR were submitted to four simulated treatments: folding, ageing with artificial saliva or sweat and washing cycles. The air permeability, mechanical integrity, electrostatic potential, and filtration efficiency (FE) of the masks were measured to quantify possible degradation. Overall, air permeability and mechanical integrity were not affected, except after washing, which slightly degraded the filtering layers. Electrostatic potential and FE showed a strong correlation, highlighting the role of electrostatic charges on small particle filtration. A slight decrease in FE for 100 nm particles was found, from 74.4% for the reference masks to 70.6% for the mask treated in saliva for 8 h. A strong effect was observed for washed masks, resulting in FE of 46.9% (± 9.5%), comparable to that of a control group with no electrostatic charges. A dry store and reuse strategy could thus be envisaged for the public if safety in terms of viral and bacterial charge is ensured, whereas washing strongly impacts FE and is not recommended.
UR - http://www.scopus.com/inward/record.url?scp=85126848911&partnerID=8YFLogxK
U2 - 10.1038/s41598-022-09068-1
DO - 10.1038/s41598-022-09068-1
M3 - Article
C2 - 35322142
AN - SCOPUS:85126848911
VL - 12
JO - Scientific Reports
JF - Scientific Reports
SN - 2045-2322
IS - 1
M1 - 4938
ER -