Abstract
Uniform large transition-edge sensor (TES) arrays are fundamental for the next generation of x-ray space observatories. These arrays are required to achieve an energy resolution ΔE < 3 eV full width at half maximum (FWHM) in the soft x-ray energy range. We are currently developing x-ray microcalorimeter arrays for use in the future laboratory and space-based x-ray astrophysics experiments and ground-based spectrometers. In this contribution, we report on the development and the characterization of a uniform 32 × 32 pixel array with 140 × 30 μm2 Ti/Au TESs with the Au x-ray absorber. We report on extensive measurements on 60 pixels in order to show the uniformity of our large TES array. The averaged critical temperature is Tc = 89.5 ± 0.5 mK, and the variation across the array (∼1 cm) is less than 1.5 mK. We found a large region of detector's bias points between 20% and 40% of the normal-state resistance where the energy resolution is constantly lower than 3 eV. In particular, results show a summed x-ray spectral resolution ΔEFWHM = 2.50 ± 0.04 eV at a photon energy of 5.9 keV, measured in a single-pixel mode using a frequency domain multiplexing readout system developed at SRON/VTT at bias frequencies ranging from 1 MHz to 5 MHz. Moreover, we compare the logarithmic resistance sensitivity with respect to temperature and current (α and β, respectively) and their correlation with the detector's noise parameter M, showing a homogeneous behavior for all the measured pixels in the array.
| Original language | English |
|---|---|
| Article number | 023101 |
| Number of pages | 10 |
| Journal | Review of Scientific Instruments |
| Volume | 92 |
| Issue number | 2 |
| DOIs | |
| Publication status | Published - 2021 |
Fingerprint
Dive into the research topics of 'Performance and uniformity of a kilo-pixel array of Ti/Au transition-edge sensor microcalorimeters'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver