Temperature and pressure effects on the performance of the portable TSI 3007 condensation particle counter: Implications on ground and aerial observations

S. Bezantakos, G. Biskos

Research output: Contribution to journalArticleScientificpeer-review

1 Citation (Scopus)
31 Downloads (Pure)

Abstract

One of the most commonly employed instruments in the category of handheld condensation particle counters (CPCs) is the TSI 3007, which employs a simplified system for achieving a temperature difference between its saturator and its condenser in view of increasing portability, energy efficiency and autonomy. As a result, particle growth and consequently the detection efficiency of the instrument can be affected by the measurement conditions. In this work we measure the detection efficiency of the TSI 3007 CPC under temperatures and pressures that differ from standard conditions. Our results show that the performance of the CPC has a strong temperature dependence in the range of 5–30 °C, whereas it is not affected by ambient pressure when this varies between 0.7 and 1.0 atm. The temperature dependent detection efficiency of the instrument becomes significant at sizes below 20 nm. Recording the temperature at which this CPC is operated is therefore strongly advised, especially when required to determine the number concentration of sub-20 nm aerosol particles.
Original languageEnglish
Article number105877
Pages (from-to)1-9
Number of pages9
JournalJournal of Aerosol Science
Volume159
DOIs
Publication statusPublished - 2022

Bibliographical note

Green Open Access added to TU Delft Institutional Repository 'You share, we take care!' - Taverne project https://www.openaccess.nl/en/you-share-we-take-care
Otherwise as indicated in the copyright section: the publisher is the copyright holder of this work and the author uses the Dutch legislation to make this work public.

Keywords

  • TSI 3007
  • Portable CPC
  • Outdoor aerosol measurements
  • UAV Aerosol measurements

Fingerprint

Dive into the research topics of 'Temperature and pressure effects on the performance of the portable TSI 3007 condensation particle counter: Implications on ground and aerial observations'. Together they form a unique fingerprint.

Cite this