Using Image Quality Metrics to Optimize the Design of Integrated Medical Ultrasound ADCs

Nikola Radeljic-Jakic*, Adriaan J. Flikweert, Nuriel N.M. Rozsa, Hendrik J. Vos, Michiel A.P. Pertijs

*Corresponding author for this work

Research output: Contribution to journalArticleScientificpeer-review

Abstract

Emerging handheld and wearable ultrasound devices enable diagnosis and long-term monitoring outside clinical settings. They require a low-power, highly complex, locally integrated system to process the RF data. The analog-to-digital converter (ADC) is a critical building block in the receive chain of these systems as it enables digital beamforming and image reconstruction. However, the ADCs currently used in cart-based imaging systems are bulky and consume too much power to be integrated into battery-powered devices. This article investigates how the area and power consumption of the commonly used successive approximation register (SAR) ADC can be reduced without negatively affecting B-mode and color-Doppler image quality. A Monte Carlo (MC) simulation study was performed in which RF data acquired with a phased-array transducer in Field II were digitized using a model of a nonideal ADC. Five different nonidealities were applied to four commonly used SAR-ADC architectures. B-mode and color-Doppler images were reconstructed from the digitized RF data. The impact of the nonidealities on the image quality was evaluated by means of three image quality metrics (IQM): peak signal-to-noise ratio (PSNR), structural similarity index (SSIM), and contrast-to-noise ratio (CNR). The effectiveness of error correction and ways of calibration are also discussed. The results show that both B-mode imaging and color-Doppler imaging are inherently resilient to nonidealities, particularly capacitor mismatch, leading to relaxed ADC requirements and paving the way for more practical in-probe digitization.

Original languageEnglish
Pages (from-to)1065-1078
Number of pages14
JournalIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control
Volume72
Issue number8
DOIs
Publication statusPublished - 2025

Bibliographical note

Green Open Access added to TU Delft Institutional Repository as part of the Taverne amendment. More information about this copyright law amendment can be found at https://www.openaccess.nl. Otherwise as indicated in the copyright section: the publisher is the copyright holder of this work and the author uses the Dutch legislation to make this work public.

Keywords

  • Application-specific integrated circuit (ASIC)
  • error correction
  • image quality metrics (IQM)
  • in-probe digitization
  • medical ultrasound
  • successive approximation analog-to-digital converter (SAR-ADC)