Three-dimensional localization microscopy with increased axial precision through TIRF angle modulation

Daniel Fan, Jelmer Cnossen, Shih Te Hung, Dimitri Kromm, Nynke H. Dekker, Gerard J. Verbiest, Carlas S. Smith*

*Corresponding author for this work

Research output: Contribution to journalArticleScientificpeer-review

33 Downloads (Pure)

Abstract

To better understand the interactions between biological molecules, a high optical resolution in all three dimensions is crucial. The intrinsically lower axial resolution of microscopes however, is a limiting factor in fluorescence imaging, correspondingly in fluorescence based single molecule localization microscopy (SMLM). Here, we present a method to improve the axial localization precision in SMLM by combining point-spread-function engineering with total internal reflection fluorescence (TIRF) fields with decay lengths that vary within the on-time of a fluorophore. Such time-varying illumination field intensity allows one to extract additional axial location information from the emitted photons. With this time varying illumination approach, we show that axial localization is improved two-fold over TIRF-based SMLM using astigmatic PSFs. We calculate theoretical resolution gains for various imaging conditions via the Cramér Rao Lower Bound (CRLB), a commonly used metric to compute the best attainable localization precision in SMLM.

Original languageEnglish
Article number129548
Number of pages7
JournalOptics Communications
Volume542
DOIs
Publication statusPublished - 2023

Keywords

  • PSF engineering
  • Single molecule localization microscopy
  • Super-resolution microscopy
  • TIRF microscopy
  • Total internal reflection fluorescence

Fingerprint

Dive into the research topics of 'Three-dimensional localization microscopy with increased axial precision through TIRF angle modulation'. Together they form a unique fingerprint.