Segmented RF shield design to minimize eddy currents for low-field Halbach MRI systems

Bart de Vos*, Rob Remis, Andrew Webb

*Corresponding author for this work

Research output: Contribution to journalArticleScientificpeer-review

9 Downloads (Pure)

Abstract

MRI systems have a thin conducting layer placed between the gradient and RF coils, this acts as a shield at the RF-frequency, minimizing noise coupled into the experiment, and decreasing the coupling between the RF and gradient coils. Ideally, this layer should be transparent to the gradient fields to reduce eddy currents. In this work the design of such a shield, specifically for low-field point-of-care Halbach based MRI devices, is discussed. A segmented double layer shield is designed and constructed based on eddy current simulations. Subsequently, the performance of the improved shield is compared to a reference shield by measuring the eddy current decay times as well as using noise measurements. A maximum reduction factor of 2.9 in the eddy current decay time is observed. The segmented shield couples in an equivalent amount of noise when compared to the unsegmented reference shield. Turbo spin echo images of a phantom and the brain of a healthy volunteer show improvements in terms of blurring using the segmented shield.

Original languageEnglish
Article number107669
Number of pages9
JournalJournal of Magnetic Resonance
Volume362
DOIs
Publication statusPublished - 2024

Keywords

  • Eddy currents
  • Halbach-array
  • Low-field
  • Shield design